VIBRATION TEST SYSTEMS * 10246 CLIPPER COVE * AURORA, OHIO 44202 * 330/562-5729


Calibration Systems / With Back-To-Back Calibration Traceable to NIST

VibratorVTS Calibrators provide an NIST traceable, back-to-back comparison calibration of accelerometers and vibration transducers. Five force ratings are available.

The calibrator uses a dual leaf spring system to reduce acceleration waveform distortion to 5% or less in the frequency range from 5 Hz-50 Hz at 1/2 inch peak-peak displacement.

The nominal distortion from 100 Hz to 10 KHz is about 1% with three narrow peaks of about 3%. Lateral acceleration is below 10% up to about 4 KHz. Frequency range is 10 Hz to 5 KHz. Modifications to extend the frequency range up to 10 KHz are available (maximum force range is decreased to 80 lbf).

Spec Sheet

Calibration
System Specification

 

VTS 40C

VTS 65C

VTS 100C

VTS 80C*

Rated Force

 

40lbf

65 lbf

100lbf

Amplifier Cooling Fan

 

Included

Included

Included

Vibrator Cooling Blower

 

Not Req'd

Required

Included

Dynamic Stroke,
peak to peak

 

0.75"

0.75"

0.75"

Stroke Limit,
between stops

 

1.0"

1.0"

1.0"

Rated Velocity

 

35 in/sec.

70 in/sec.

100 in/sec.

Maximum acceleration,
w/cal. std.

 

40g
32g* w/al armature

65g
52g* w/al armature

100g
80g* w/al armature

Armature Weight,
w/cal std.

 

1.0 lbs

1.0 lbs

1.0 lbs

Suspension Stiffness

 

50 lbs/in.

50 lbs/in.

50 lbs/in.

First Major Resonance,
w/cal. std.

 

4200 Hz

4200 Hz

4200 Hz

Useful Frequency Range,

w/cal. std.

10-5000 Hz

10-5000 Hz

10-5000 Hz

w/cal. std.
& aluminum armature*

10 Hz-10 KHz*

10 Hz-10 KHz*

10 Hz-10 KHz*

System Power Required
100-240 V single phase 50/60 Hz

 

3a/120V
1.5a/240V

10a/120V
5a/240V

20a/120V
10a/240V

Specifications subject to change without notice.


VTS